Load samples into the sample
cups
Follow the instructions below to load the samples into the sample cups.
Action
Step
A maximum of 150 μl of sample may be placed in these cups.
1
Check to make sure that there is cover fluid over the samples. When the
sample is introduced into the cups, it will sink through the oil to the bottom
of the cup and come into contact with the IPG strip.
2
Close the Ettan IPGphor 3 lid.
3
4
•
Program and run the Ettan IPGphor 3 from its own control panel display.
See Section 5.3 Perform a run from the control panel, on page 68 for more
details, or
•
Start the Ettan IPGphor 3 from the Ettan IPGphor 3 Control Software.
See Section 5.4.1 Setting up a run, on page 91 for more details.