5.1.2
Use the Ettan IPGphor Cup Loading Manifold
Precautions
CAUTION
Always wear protective gloves when working with rehydration so-
lution.
Introduction
For more detailed information, please refer to Ettan IPGphor Cup Loading Manifold User
Manual.
Rehydrate Immobiline DryStrip
in the IPGbox
Rehydrate the IPG strips with the gel side down in the appropriate volume of rehydration
solution, with or without sample, using the IPGbox.
Rehydration in the Cup Loading Manifold is not recommended, the channel is too wide
to ensure proper rehydration.
Also, follow the instructions included with the Immobiline DryStrips or the IPGbox. Close
the lid of the IPGbox and allow the strips to rehydrate overnight (10 to 20 hours).
Ettan IPGphor 3 Operating Instructions 28-9635-51 AC
59
5 Operation
5.1 Preparations before start
5.1.2 Use the Ettan IPGphor Cup Loading Manifold