Setup
Place the instrument
WARNING
The vents at the front, back, and bottom of Ettan IPGphor 3 must
not be obstructed. Air must be free to circulate for the cooling ap-
paratus to function correctly. There should be a minimum of 10
cm unobstructed space behind the rear panel of the instrument.
Action
Step
Select a place for Ettan IPGphor 3 where it is possible to access the mains
power switch on the rear panel of the instrument.
1
Position the instrument on a flat surface with the spirit level in the center.
Turn the leveling feet as necessary to make the instrument level and stable.
2
Power connection
WARNING
Make sure that the supply voltage at the wall outlet corresponds
to the marking on the instrument, before connecting the power
cord.
WARNING
Only use power cords delivered or approved by GE.
WARNING
The Ettan IPGphor 3 must always be connected to a grounded
power outlet.
Ettan IPGphor 3 Operating Instructions 28-9635-51 AC
41
4 Installation
4.3 Setup