6.1
Cleaning Ettan IPGphor 3
Precautions
WARNING
If large volumes of liquid have penetrated the casing of the instru-
ment and come into contact with the electrical components, imme-
diately switch off the instrument and contact an authorized service
technician.
NOTICE
Do not use strong acids, bases, ketones, alcohols, or other reagents
to clean the covers, sample cups, or electrodes or the parts may
be damaged.
Cups may be briefly rinsed with ethanol if desired.
NOTICE
Clean strip holders and manifold with the strip holder cleaning so-
lution provided. Clean all other components that come in contact
with the IPG strip or the sample with a detergent designed for
glassware. Rinse well with distilled water.
Clean Ettan IPGphor 3
The parts of the instrument that come in contact with liquid reagents are resistant to
chemicals typically used for IEF. Before introducing any other chemicals into the system,
first test the affected parts.
Cleaning procedures
IPGphor Strip Holder Cleaning Solution has been specifically formulated for removing
protein deposits. It will not damage the strip holder or manifold. IPGphor Strip Holder
Cleaning Solution (29-0119-27) can be ordered in 1 liters bottles from GE.