- 90 -
eBike III basic / comfort
2018111-336 Rev. B
o
pen
the
electronic
carrier
houSing
(
bottom
Side
)
I
nstallatIon
of
tHer
CoM - I
nterfaCe
extensIon
The MAC 5000 ST / MAC 5500 use analogue signals
to set the load of the eBike, while the bloodpressure
measurement is controlled with digital signals..
The eBike III is by default only equipped with a digital
interface.
To connect an eBike III with a MAC 5000 ST / 5500 a
special adapter (COM module PN 2018111-340) has to be
installed to add an analogue interface to the eBike III.
A jumper adapter has to be placed onto the LIU board to
connect all required signals to the DSUB connector:
1. Open the electronic carrier housing located in the
bottom plate of the eBike III of the BKE on the bottom
side of the ergometer.
com
module
with
cable
,
jumper
and
velcro
tape