2018111-336 Rev. B
eBike III basic / comfort
- 73 -
Service Menu
Contrast
ECG Type
Baudrate 8N2
Analog Calibration
Software update
Dynamic Load Calibr.
Saddle Calibration
Dynamic Load Cal.
Calibration offset
Gain: 1,00
DMS:
3277 / 3219
RPM:
0 B
0 P
Load:
0 SW
0 AW
START
Dynamic Load Cal.
Calibration offset
Gain: 1,00
DMS:
3277 / 3222
RPM:
0 B
0 P
Load:
0 SW
0 AW
WAITING...
Dynamic Load Cal.
Dyn. Calibration
Gain: 1,00
DMS:
3219 / 3219
RPM:
0 B
0 P
Load:
0 SW
AW
START
Dynamic Load Cal.
At 50
/50 W 51,3 W
Gain: 1,00
DMS:
3219 / 3219
RPM:
50 B
0 P
Load:
50 SW
AW
SET
d
ynamic
load
offSet
calibration
-
waiting
for
Start
Select „Dynamic load calibration“ and confirm with NIBP
• Start the offset calibration process with NIBP
• The line „DMS“ shows:
stored DMS value / actual DMS value
• By pressing NIBP (START) the actual value is averaged
for 5 sec and stored
• Press NIBP (START) to start calibration step #1
• The eBike III automatically sets the load to 50 Watt
• Set the calibrator unit to 50 rpm
• The line „RPM“ shows the measured rotations/min
(issued by the calibrator unit)
[the value „P
“ is not used in the eBike III)
• The line „Load“ shows the target load in Watt-
automatically set by the eBike III (e.g. 50 Watt)
[the value „AW“ is not used in the eBike III)
• Wait approx. 20 seconds, until the load display is
stable
• By using the arrow keys, input the actual measured
load value of the calibrator (e.g. 51,3 Watt)
• Store the value with SET.
S
election
of
parameterS
in
Service
menu
d
ynamic
load
calibration
Step
1 -
waiting
for
Start
d
ynamic
load
offSet
calibration
-
averaging
d
ynamic
load
calibration
Step
1 -
input
load
value