- 92 -
eBike III basic / comfort
2018111-336 Rev. B
Service Code
999
Service Menu
Saddle calibration
NIBP calibration
NIBP test
System config.
Error log
Manual load
Reset DMS values
System Config.
Saddle engine
0 I
NIBP-Modul
0 I
Analog Interface
0 I
EXIT
e
B
Ike
III
software
settIng
To activate the service menu, the ergometer has to be
powered off and on again.
As soon as the self test displays on the ergometer LCD,
press the two arrow keys simultaneously:
To prevent settings from being changed in error, the
service menu is “protected“ with a password.
Select “999“ with the arrow keys and confirm by pressing
NIBP.
Select “System config.“ by using the arrow keys and
confirm by pressing NIBP
Select “Analog Interface“ by using the arrow keys and
change the setting to “I“ by pressing NIBP.
Then select “EXIT“ and return to the service menu.by
pressing NIBP.
Then power OFF/ON the eBike.
Note
• The “ECG type” can not be changed until the new system
configuration has been activated by powering the eBike
III OFF and ON again