2018111-336 Rev. B
eBike III basic / comfort
- 123 -
t
eCHnICal
s
peCIfICatIons
e
B
Ike
III B
lood
pressure
Module
Measuring Method:
auscultatory method, oscillometric; for resting BP, the
results from both measurements are compared for
plausibility
Measuring Range
systolic pressure:
40 to 280 mmHg
diastolic pressure: 40 to 280 mmHg
pulse rate:
35 to 230 P/min
Measurement Error
pressure readout error: +/- 3 mmHg
readout resolution:
+/- 1 mmHg
Inflation Pressure
300 mmHg max.; during inflation the inflation pressure
automatically adapts to patient's BP
Inflation Rate
between approx. 6 seconds (to 140 mmHg) and
approx. 18 seconds (to 300 mmHg)
Max. Cuff Pressure
300 mmHg
Cuff Deflation Method
pulse-dependent deflation rate
approx. 3 mmHg/beat or approx. 3 mmHg/s
Calibration
calibration with external pressure meter
Artifact Rejection
automatic artifact rejection and comparison of the
resting BP readings from both methods for plausibility