Service Menu
Contrast
ECG type
NIBP mmHg / kPa
NIBP Beep
Baudrate 8N2
Analog Calibration
Software update
Service Code
999
s
ervICe
Menu
To activate the service menu, the ergometer has to be
powered off and on again.
As soon as the self test displays on the ergometer LCD,
press the two arrow keys simultaneously:
To prevent settings from being changed in error, the
service menu is “protected“ with a password.
Select “999“ with the arrow keys and confirm by pressing
NIBP.
The service menu with the following parameters displays:
•
Contrast
•
ECG type
•
NIBP mmHg / kPa
•
NIBP Beep
•
Baudrate 8N2
•
Analog Calibration
•
Software Update
•
Dynamic Load Calibration
•
Saddle Calibration
•
NIBP Calibration
•
NIBP Test
•
System Config.
•
Error log
•
Manual load
•
Reset DMS values
Navigate through the menus by using the arrow keys
(saddle height) to position the cursor bar.
Confirm the selection by pressing NIBP (to the right
of the LCD).
Select the desired parameter using the arrow keys and
confirm by pressing NIBP.
Note
• The service menu in the ergometers is available in
english only
• Depending on the actual ergometer configuration,
not all settings are enabled
• Leaving the service menu is only possible by powering the
device off
• The following descriptions are not in the same order as in
the service menu
p
aSSword
for
Service
menu
S
election
of
parameterS
in
Service
menu