- 72 -
eBike II, eBike II L, eBike II EL
2018112-069 Rev. F
All the different components are connected with tubes
(one pressure system) and connected to the cuff connector
of the ergometer control terminal (use cuff adaptor).
nIBp t
est
This function is used to check the internal pressure system
(pressure sensor, closeness, leakage etc.).
Select „NIBP Test“ in the service menu and confirm with
ENTER (or SELECT).
Using the hand pump, apply a fixed pressure to the system
The actual pressure value is displayed.
Compare the displayed ergometer value with the value of
the manometer.
The maximum allowed deviation is +/- 3mmHg.
compensating container
manometer
hand pump
c
alibration
Setup
nibp
t
eSt
of
the
internal
nibp
preSSure
SySteM
NIBP Test
NIBP pressure
123 mmHg
CLOSE VALVE
Hint
• If the applied pressure reaches 320 mmHg, the internal
safety valve is opened.
Press „Close valve“ (middle soft key) to continue with the
NIBP-Test.
Summary of Contents for eBike II Series
Page 1: ...GE Healthcare eBike II eBike II L eBike II EL Service Manual 2018112 069 Revision F...
Page 2: ......
Page 13: ...2018112 069 Rev F eBike II eBike II L eBike II EL 13...
Page 39: ...2018112 069 Rev F eBike II eBike II L eBike II EL 39...
Page 47: ...2018112 069 Rev F eBike II eBike II L eBike II EL 47...
Page 77: ...2018112 069 Rev F eBike II eBike II L eBike II EL 77...
Page 133: ...2018112 069 Rev F eBike II eBike II L eBike II EL 133...
Page 141: ...2018112 069 Rev F eBike II eBike II L eBike II EL 141...
Page 212: ...212 eBike II eBike II L eBike II EL 2018112 069 Rev F...
Page 214: ...214 eBike II eBike II L eBike II EL 2018112 069 Rev E...
Page 216: ...216 eBike II eBike II L eBike II EL 2018112 069 Rev E 210 eBike II L EL uncrating instructions...
Page 228: ...228 eBike II eBike II L eBike II EL 2018112 069 Rev E 222 eBike II L EL uncrating instructions...
Page 233: ......