2018112-069 Rev. F
eBike II, eBike II L, eBike II EL
- 45 -
dMs
exCHange
• Remove the DMS cable from the LRE board
• Loose 2 Allen key screws
• The strain gauge unit hangs at the tensioning rope
• Scratch the sealing wax from the top suspension of
the tensioning rope (e.g. by use of a screw driver)
• Remove the tensioning rope to the top by pulling it off
the vertical slot
The assembly of the replacement DMS unit is performed in
reverse order.
WARNING: The rubber coated side of the strain
gauge should not be hurt !!
To fix the tensioning rope in the upper fixture, a standard
glue can be used.