- 210 -
eBike II, eBike II L, eBike II EL
2018112-069 Rev. F
t
eCHnICal
s
peCIfICatIons
e
B
Ike
II B
lood
p
ressure
M
odule
Measuring Method:
auscultatory method, oscillometric; for resting BP, the
results from both measurements are compared for
plausibility
Measuring Range
systolic pressure:
40 to 280 mmHg
diastolic pressure: 40 to 280 mmHg
pulse rate:
35 to 230 P/min
Measurement Error
pressure readout error: +/- 3 mmHg
readout resolution:
+/- 1 mmHg
Inflation Pressure
300 mmHg max.; during inflation the inflation pressure
automatically adapts to patient's BP
Inflation Rate
between approx. 6 seconds (to 140 mmHg) and
approx. 18 seconds (to 300 mmHg)
Max. Cuff Pressure
300 mmHg
Cuff Deflation Method
pulse-dependent deflation rate
approx. 3 mmHg/beat or approx. 3 mmHg/s
Calibration
calibration with external pressure meter
Artifact Rejection
automatic artifact rejection and comparison of the
resting BP readings from both methods for plausibility
Summary of Contents for eBike II Series
Page 1: ...GE Healthcare eBike II eBike II L eBike II EL Service Manual 2018112 069 Revision F...
Page 2: ......
Page 13: ...2018112 069 Rev F eBike II eBike II L eBike II EL 13...
Page 39: ...2018112 069 Rev F eBike II eBike II L eBike II EL 39...
Page 47: ...2018112 069 Rev F eBike II eBike II L eBike II EL 47...
Page 77: ...2018112 069 Rev F eBike II eBike II L eBike II EL 77...
Page 133: ...2018112 069 Rev F eBike II eBike II L eBike II EL 133...
Page 141: ...2018112 069 Rev F eBike II eBike II L eBike II EL 141...
Page 212: ...212 eBike II eBike II L eBike II EL 2018112 069 Rev F...
Page 214: ...214 eBike II eBike II L eBike II EL 2018112 069 Rev E...
Page 216: ...216 eBike II eBike II L eBike II EL 2018112 069 Rev E 210 eBike II L EL uncrating instructions...
Page 228: ...228 eBike II eBike II L eBike II EL 2018112 069 Rev E 222 eBike II L EL uncrating instructions...
Page 233: ......