- 70 -
eBike II, eBike II L, eBike II EL
2018112-069 Rev. F
• Set the calibrator unit to 110 rpm.
• Wait approx. 20 seconds, until the load display is
stable
• By using the arrow keys, input the actual value of the
calibrator (e.g. 55,0 Watt) at the ergometer.
• Store the value with SET.
• Set the ergometer to 400 Watt (via PC interface)
• Wait approx. 20 seconds, until the load display is
stable
• By using the arrow keys, input the actual value of the
calibrator (e.g. 412,5 Watt) at the ergometer.
• Store the value with SET.
The successfull completion of the dynamic load calibration
is confirmed.
• Stop the calibration unit.
d
ynaMic
load
calibration
- 50 w
att
at
110
rpM
Dynamic Load Cal.
Set at 110 rpm / 50 Watt
Rotation: 50
DMS-Volt: 327
mV
LRE Load:
50
Watt
Dyn. Load
55,0
Watt
SET
d
ynaMic
load
calibration
- 400 w
att
at
110
rpM
Dynamic Load Cal.
Set at 110 rpm / 400 Watt
Rotation: 50
DMS-Volt: 327
mV
LRE Load:
400
Watt
Dyn. Load 412,5
Watt
SET
d
ynaMic
load
calibration
-
finiShed
Dynamic Load Cal.
Rotation: 110
DMS-Volt: 1327
mV
LRE Load:
400
Watt
Calibration successfull
Summary of Contents for eBike II Series
Page 1: ...GE Healthcare eBike II eBike II L eBike II EL Service Manual 2018112 069 Revision F...
Page 2: ......
Page 13: ...2018112 069 Rev F eBike II eBike II L eBike II EL 13...
Page 39: ...2018112 069 Rev F eBike II eBike II L eBike II EL 39...
Page 47: ...2018112 069 Rev F eBike II eBike II L eBike II EL 47...
Page 77: ...2018112 069 Rev F eBike II eBike II L eBike II EL 77...
Page 133: ...2018112 069 Rev F eBike II eBike II L eBike II EL 133...
Page 141: ...2018112 069 Rev F eBike II eBike II L eBike II EL 141...
Page 212: ...212 eBike II eBike II L eBike II EL 2018112 069 Rev F...
Page 214: ...214 eBike II eBike II L eBike II EL 2018112 069 Rev E...
Page 216: ...216 eBike II eBike II L eBike II EL 2018112 069 Rev E 210 eBike II L EL uncrating instructions...
Page 228: ...228 eBike II eBike II L eBike II EL 2018112 069 Rev E 222 eBike II L EL uncrating instructions...
Page 233: ......