2018112-069 Rev. F
eBike II, eBike II L, eBike II EL
- 137 -
e
rror
Code
BIts
Mae
Bit
Error
Reaction
Display / Info
Troubleshooting
0
Error seat sensor
Module is switched off
Internal error
- check cable sensor <-> MAE - -
- check cable motor <-> MAE
- if error persists:
replace seat motor
1
Seat no movement
Module is switched off
Internal error
- check cable sensor <-> MAE - -
- check cable motor <-> MAE
- if error persists:
replace seat motor
2
Wrong cabling
Module is switched off
Internal error
- check cable sensor <-> MAE
- check cable motor <-> MAE
- replace cables
3
not used
4
+24V out of range
Warning
- check power supply
- check connections:
power supply to LRE
cable LRE <-> MAE
- replace MAE
5
not used
6
not used
7
Safety state mode
Module is switched off
Internal error
- restart eBike (OFF / ON)
- if error persists: replace MAE
8
not used
9
CAN zykl. Timeout
Module is switched off
Internal error
- restart eBike (OFF / ON)
- if error persists:
check cable LRE <-> MAE
- replace MAE
Summary of Contents for eBike II Series
Page 1: ...GE Healthcare eBike II eBike II L eBike II EL Service Manual 2018112 069 Revision F...
Page 2: ......
Page 13: ...2018112 069 Rev F eBike II eBike II L eBike II EL 13...
Page 39: ...2018112 069 Rev F eBike II eBike II L eBike II EL 39...
Page 47: ...2018112 069 Rev F eBike II eBike II L eBike II EL 47...
Page 77: ...2018112 069 Rev F eBike II eBike II L eBike II EL 77...
Page 133: ...2018112 069 Rev F eBike II eBike II L eBike II EL 133...
Page 141: ...2018112 069 Rev F eBike II eBike II L eBike II EL 141...
Page 212: ...212 eBike II eBike II L eBike II EL 2018112 069 Rev F...
Page 214: ...214 eBike II eBike II L eBike II EL 2018112 069 Rev E...
Page 216: ...216 eBike II eBike II L eBike II EL 2018112 069 Rev E 210 eBike II L EL uncrating instructions...
Page 228: ...228 eBike II eBike II L eBike II EL 2018112 069 Rev E 222 eBike II L EL uncrating instructions...
Page 233: ......