4 Command and Query Summary
K0472 Revision A
4-46
:SENS:MACH:INL:TIME
Applicability:
PACE6000
Command Syntax
:SENSe[x][:MACH]:INLimits:TIME <number>
where x = 1 or 2 is the module number (default - 1)
Parameter:
<integer> MACH in-limits time in seconds
Short form:
:SENS:MACH:INL:TIME <number>
Function:
Sets the time that the MACH value has to be within limits before
generating a service request.
Default:
2 seconds
minimum 2 seconds
maximum 999 seconds
Query Syntax
:SENSe[x][:MACH]:INLimits:TIME?
where x = 1 or 2 is the module number (default - 1)
Short form:
:SENS:MACH:INL:TIME?
Function:
Query MACH in-limits timers.
Response:
Number representing MACH in-limits time in seconds.
Description
Sets the MACH in-limits timer value.
Note: If the instrument is not in aero mode it returns an error: -221. Settings conflict.