PACE Series SCPI Manual
K0472 Revision A
4-7
:CAL:PRES:ZERO:VALV
Caution:
Opening the zero valve with high pressure in the system can cause damage to the
equipment. Reduce the system pressure and make sure the controller is OFF
before opening the zero valve.
Applicability:
PACE5000, PACE6000
Command Syntax
:CALibration:[PRESsure]:ZERO:VALVe[STATe]
where: x (module) = 1 or 2 (default - 1)
Parameter:
<boolean>
Short form:
:CAL:ZERO:VALV
Function:
Opens and closes zero valve.
Default:
0
Query Syntax
:CALibration:[PRESsure]:ZERO:VALVe[STATe]?
where: x (module) = 1 or 2 (default - 1)
Short form:
CAL:ZERO:VALV?
Function:
Queries state of valve.
Response:
1
-
open
0
-
close
Description
This command is used to open and close the zero valve. The query gets the state of the zero
valve - open or close.
Example
TX> :CAL:PRES:ZERO:VALV 1
TX> :CAL:PRES:ZERO:VALV:STAT 1
Either of above two commands opens the zero valve.
Either: TX> :CAL:PRES:ZERO:VALV?
RX> :CAL:PRES:ZERO:VALV 1
Or TX> :CAL:PRES:ZERO:VALV:STAT?
RX> :CAL:PRES:ZERO:VALV:STAT 1
TX> :CAL:PRES:ZERO:VALV: 0
TX> :CAL:PRES:ZERO:VALV:STAT 0
Either of above two commands close the zero valve,
switch OFF the controller.
Either: TX> :CAL:PRES:ZERO:VALV?
RX> :CAL:PRES:ZERO:VALV 0
Or TX> :CAL:PRES:ZERO:VALV:STAT?
RX> :CAL:PRES:ZERO:VALV:STAT 0
Gets the condition of the zero valve.