4 Command and Query Summary
K0472 Revision A
4-102
:STAT:OPER:PRES:ENAB
Applicability:
PACE1000, PACE5000, PACE6000
Command Syntax
:STATus:OPERation:PRESsure:ENABle <integer>
Parameter:
<integer> 16 bit value to set pressure enable bits
Short form:
STAT:OPER:PRES:ENAB <integer>
Function:
Controls the pressure status operation enable register.
Default:
0
minimum 0
maximum 32767
Query Syntax
:STATus:OPERation:PRESsure:ENABle?
Short form:
:STAT:OPER:PRES:ENAB?
Function:
Query enable register.
Response:
16 bit value of pressure enable register.
Description
Controls the bits that pass through the status reporting system, see status reporting section.