PACE Series SCPI Manual
K0472 Revision A
5 - 1
5
ERRORS
Negative error numbers are used for standard SCPI errors. Positive error numbers are device
specific errors. Following the error number, a message describes the error. An error, when
detected, is held in the error queue.
When SYST:ERR? is sent any error in the error queue sets the error bit in the event status
register.
Error numbers
Each error detected causes an error number with an error message to be returned as
follows:
Table 5-1 Errors -100 to -199
Error
code
Error message
Error
code
Error message
-102 -102, “Syntax error”
-128 -128, “Numeric data not allowed”
-104 -104, “Data type error”
-130 -130, “Suffix error”
-108 -108, “Parameter not allowed”
-131 -131, “Invalid suffix”
-109 -109, “Missing parameter”
-134 -134, “Suffix too long”
-110 -110, “Command Header Error”
-138 -138, “Suffix not allowed”
-111 -111, “Header Separator Error”
-140 -140, “Character data error”
-112 -112, “Program mnemonic too long”
-141 -141, “Invalid character data”
-113 -113, “Undefined header”
-144 -144, “Character data too long”
-114 -114, “Header suffix out of range”
-148 -148, “Character data not allowed”
-120 -120, “Numeric data error”
-150 -150, “String data error”
-121 -121, “Invalid character in number”
-151 -151, “Invalid string data”
-123 -123, “Exponent too large”
-158 -158, “String data not allowed”
-124 -124, “Too many digits”