Calibration procedures 10-1
Chapter 10: Calibration procedures
10.1 Introduction
Note:
GE can provide a calibration service that is traceable to
international standards.
Return of the instrument to the manufacturer (recommended)
or an approved service agent for calibration.
When using an alternative calibration facility, make sure that it
uses these standards.
10.2 Before starting
To do an accurate calibration:
•
use the calibration equipment specified in
Table 10-1
.
•
calibrate in a stable temperature environment: 21 ± 1°C (70
± 2°F).
Note:
Before starting a calibration procedure, leave the
instrument in the calibration environment for a minimum
period of two hours.
Table 10-1:
Calibration equipment
Function
Calibration equipment
(ppm = parts per million)
Current
(CH1 or CH2)
Current (mA) calibrator.
Accuracy - Current measure/source; see
Table 10-2
/
10-3
.
Voltage
(CH1 or CH2)
Volts calibrator.
Accuracy - Voltage measure/source; see
Table 10-5
/
10-7
.
Millivolts
(CH1 or CH2)
OR
TC mV (CH1)
mV calibrator.
Accuracy - Millivolts measure/source; see
Table 10-4
/
10-6
.
Accuracy - TC mV; see
Table 10-12
.
Frequency
(CH1)
(measure)
(source)
Signal generator
Total error: 0.3 ppm or better
Frequency meter
Total error: 0.3 ppm or better
Resolution: 8 digits (minimum)
Amplitude (source)
Digital voltmeter (DVM)
Summary of Contents for Druck DPI 620-IS
Page 4: ...Page 4 of 98 K0460 Issue 1 ...
Page 5: ...GE Measurement Control Systems Druck DPI 620 IS advanced modular calibrator user manual K0460 ...
Page 12: ...Issue 1 viii About this manual K0460 EN English ...
Page 21: ... EN English K0460 Instrument parts accessories and options 1 5 Issue 1 ...
Page 22: ...Issue 1 1 6 Instrument parts accessories and options K0460 EN English ...
Page 42: ...Issue 1 2 20 Prepare the instrument K0460 EN English ...
Page 60: ...Issue 1 5 2 Instrument communications K0460 EN English ...
Page 64: ...Issue 1 6 4 Datalog operation K0460 EN English ...
Page 78: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Page 94: ...Issue 1 10 16 Calibration procedures K0460 EN English ...
Page 96: ...Issue 1 11 2 General specification K0460 EN English ...
Page 97: ...Customer service Visit our web site www gesensinginspection com ...
Page 98: ......