[EN] English - K0460
Prepare the instrument 2-9
Issue 1
2.8.2 Procedure to see the instrument status
2.8.3 Procedures to make Task selections
When first using the calibrator, there are default measure
and source
functions on the display:
•
CH1 settings: RTD source, RTD type is PT100, scale is °C;
Automation is
Nudge
; see
Chapter 3
.
•
CH2 settings: Current measure, see
Chapter 3
.
Procedure Overview
Use the
Task
menu to complete these procedures:
•
Set the calibrator functions on the display: item
; see
Section 2.8.4
.
a. CH1: Channel 1 electrical function: Choose one option:
b. CH2: Channel 2 electrical function: Choose one option:
TAP
Configuration
2
1
3
Status
The amount of memory available (in MB and as a
% of the total on the device): SD Card (internal).
Charging or battery-powered, and the amount of
charge in 1% increments.
The date of the next calibration.
CH1:
source
CH2:
measure
TAP
2
1
C
B
A
None
Resistance
Current
RTD
Voltage
CJ
Millivolts
Observed*
TC (°C/°F)
TC (mV)
Frequency
None
Voltage
Current
Millivolts
Current (24V)
Observed*
A
Summary of Contents for Druck DPI 620-IS
Page 4: ...Page 4 of 98 K0460 Issue 1 ...
Page 5: ...GE Measurement Control Systems Druck DPI 620 IS advanced modular calibrator user manual K0460 ...
Page 12: ...Issue 1 viii About this manual K0460 EN English ...
Page 21: ... EN English K0460 Instrument parts accessories and options 1 5 Issue 1 ...
Page 22: ...Issue 1 1 6 Instrument parts accessories and options K0460 EN English ...
Page 42: ...Issue 1 2 20 Prepare the instrument K0460 EN English ...
Page 60: ...Issue 1 5 2 Instrument communications K0460 EN English ...
Page 64: ...Issue 1 6 4 Datalog operation K0460 EN English ...
Page 78: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Page 94: ...Issue 1 10 16 Calibration procedures K0460 EN English ...
Page 96: ...Issue 1 11 2 General specification K0460 EN English ...
Page 97: ...Customer service Visit our web site www gesensinginspection com ...
Page 98: ......