[EN] English - K0460
HART
®
device operations 8-5
Issue 1
Self-test
•
To confirm that the transmitter is functioning
correctly, navigate through the menus to the
self-test method button:
•
Press the self-test button.
•
Press OK and the self-test executes.
L
o
o
p
t
e
s
t
.
•
To confirm that the transmitter is functioning
correctly navigate through the menus to the
loop test button:
•
Press the loop test button.
•
Press OK and the loop test executes.
•
The loop test can be put into various modes,
typically 4 mA and 20 mA should be selected.
The Analogue value of Primary Variable mA
current (actual calibrated value) of the
transmitter can be seen by selecting the CH2
drop-down box. If this value needs adjusting
then use the analogue trim (sometimes referred
to as D/A trim) in the calibration menu.
Analog trim (D/A trim)
•
Using the DPI 620-IS, an analogue trim can be performed on
the 4 to 20 mA loop without having to connect any external
reference meters.
•
Navigate to the calibration folder and select the D/A trim
method. Follow the on-screen instructions.
Summary of Contents for Druck DPI 620-IS
Page 4: ...Page 4 of 98 K0460 Issue 1 ...
Page 5: ...GE Measurement Control Systems Druck DPI 620 IS advanced modular calibrator user manual K0460 ...
Page 12: ...Issue 1 viii About this manual K0460 EN English ...
Page 21: ... EN English K0460 Instrument parts accessories and options 1 5 Issue 1 ...
Page 22: ...Issue 1 1 6 Instrument parts accessories and options K0460 EN English ...
Page 42: ...Issue 1 2 20 Prepare the instrument K0460 EN English ...
Page 60: ...Issue 1 5 2 Instrument communications K0460 EN English ...
Page 64: ...Issue 1 6 4 Datalog operation K0460 EN English ...
Page 78: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Page 94: ...Issue 1 10 16 Calibration procedures K0460 EN English ...
Page 96: ...Issue 1 11 2 General specification K0460 EN English ...
Page 97: ...Customer service Visit our web site www gesensinginspection com ...
Page 98: ......