4-6 Pressure indicator operation (MC 620-IS)
K0460 - [EN] English
1.
Assemble the pressure indicator with the correct PM 620-IS
modules; see
Section 4.2.1
.
2.
Set the applicable software options; see
Section 4.4.1
(Procedure overview)
. This example shows two pressure
functions:
•
Pressure functions P1 and P2 are set-up.
3.
To attach the external equipment, see
Section 4.3.1
.
4.5 Error indications
If the display shows <<<< (under range) or >>>> (over range):
•
Make sure that the range is correct.
•
Make sure that all the related equipment and connections
are serviceable.
<<<<<
Under range: The display shows this symbol for this condition:
Reading < Negative FS - (10% of negative full-scale).
>>>>>
Over range: The display shows this symbol for this condition:
Reading > Positive FS + (10% of positive full-scale).
Summary of Contents for Druck DPI 620-IS
Page 4: ...Page 4 of 98 K0460 Issue 1 ...
Page 5: ...GE Measurement Control Systems Druck DPI 620 IS advanced modular calibrator user manual K0460 ...
Page 12: ...Issue 1 viii About this manual K0460 EN English ...
Page 21: ... EN English K0460 Instrument parts accessories and options 1 5 Issue 1 ...
Page 22: ...Issue 1 1 6 Instrument parts accessories and options K0460 EN English ...
Page 42: ...Issue 1 2 20 Prepare the instrument K0460 EN English ...
Page 60: ...Issue 1 5 2 Instrument communications K0460 EN English ...
Page 64: ...Issue 1 6 4 Datalog operation K0460 EN English ...
Page 78: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Page 94: ...Issue 1 10 16 Calibration procedures K0460 EN English ...
Page 96: ...Issue 1 11 2 General specification K0460 EN English ...
Page 97: ...Customer service Visit our web site www gesensinginspection com ...
Page 98: ......