Issue 1
3-6 Electrical operations
K0460 - [EN] English
3.2.7 Example procedure: Measure or simulate an RTD (or Resistance)
These examples (A and B) show Channel 1 (CH1) set-up to
measure or simulate an RTD. A 4-Wire configuration gives the
best accuracy; a 2-Wire configuration has the lowest accuracy
(4-Wire RTD shown).
Note:
To measure or simulate resistance (
Ω
), set the
Resistance function.
Example A
1.
Set the applicable software options; see
Section 3.2.1
(Procedure overview)
.
2.
Complete the electrical connections.
3.
If necessary change the
Settings
and continue with the
measure operation.
Values to set
:
RTD Type (Set the applicable RTD); see
Table A1
(front
cover) for the available options.
Example B
1.
Set the applicable software options; see
Section 3.2.1
(Procedure overview)
.
2.
Complete the electrical connections.
2
1
Measure an RTD on
channel 1 (CH1)
Range: 850 °C
RTD Type: PT100
Connection: 4 wire
A
3
Simulate an RTD on
channel 1 (CH1)
Range: 850 °C
RTD Type: PT100
Connection: 4 wire
Automation
:
Nudge
(
Section 2.9.2
)
2
B
4
1
3
Summary of Contents for Druck DPI 620-IS
Page 4: ...Page 4 of 98 K0460 Issue 1 ...
Page 5: ...GE Measurement Control Systems Druck DPI 620 IS advanced modular calibrator user manual K0460 ...
Page 12: ...Issue 1 viii About this manual K0460 EN English ...
Page 21: ... EN English K0460 Instrument parts accessories and options 1 5 Issue 1 ...
Page 22: ...Issue 1 1 6 Instrument parts accessories and options K0460 EN English ...
Page 42: ...Issue 1 2 20 Prepare the instrument K0460 EN English ...
Page 60: ...Issue 1 5 2 Instrument communications K0460 EN English ...
Page 64: ...Issue 1 6 4 Datalog operation K0460 EN English ...
Page 78: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Page 94: ...Issue 1 10 16 Calibration procedures K0460 EN English ...
Page 96: ...Issue 1 11 2 General specification K0460 EN English ...
Page 97: ...Customer service Visit our web site www gesensinginspection com ...
Page 98: ......