[EN] English - K0460
Calibration procedures 10-11
Issue 1
10.11 Procedures
(CH1): Resistance
(source)
1.
Connect the applicable calibration equipment in
Table 10-1
;
for example: see
Section 3.2.9 (Example procedure:
Measure or simulate an RTD (or Resistance))
, example
B
.
2.
Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
3.
Use the calibration menu (
Section 10.3
) to do a two-point
calibration for each range.
•
Range: 0-400
Ω
•
Range: 400
Ω
-2000
Ω
•
Range: 2k
Ω
-4k
Ω
4.
To make sure the calibration is correct, select the applicable
Resistance (source) function; see
Section 3.2.9
, (
B
).
5.
Set the specified values (
Table 10-11
). Make sure the error
is in the specified limits.
Table 10-11:
Resistance (source) error limits
Ohms
(
Ω
)
Excitation
(mA)
Calibrator
uncertainty (
Ω
)
Permitted
DPI 620-IS error
(
Ω
)
0
0.1
0.0014
0.014
100
0.1
0.0016
0.038
200
0.1
0.0021
0.062
300
0.1
0.0028
0.11
400
0.1
0.0035
0.31
1000
0.1
0.008
0.55
2000
0.1
0.016
0.86
4000
0.1
0.032
1.1
Summary of Contents for Druck DPI 620-IS
Page 4: ...Page 4 of 98 K0460 Issue 1 ...
Page 5: ...GE Measurement Control Systems Druck DPI 620 IS advanced modular calibrator user manual K0460 ...
Page 12: ...Issue 1 viii About this manual K0460 EN English ...
Page 21: ... EN English K0460 Instrument parts accessories and options 1 5 Issue 1 ...
Page 22: ...Issue 1 1 6 Instrument parts accessories and options K0460 EN English ...
Page 42: ...Issue 1 2 20 Prepare the instrument K0460 EN English ...
Page 60: ...Issue 1 5 2 Instrument communications K0460 EN English ...
Page 64: ...Issue 1 6 4 Datalog operation K0460 EN English ...
Page 78: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Page 94: ...Issue 1 10 16 Calibration procedures K0460 EN English ...
Page 96: ...Issue 1 11 2 General specification K0460 EN English ...
Page 97: ...Customer service Visit our web site www gesensinginspection com ...
Page 98: ......