10-2 Calibration procedures
K0460 - [EN] English
Before starting the calibration, make sure that the time and
date on the instrument are correct; see
Section 2.8.1
(Procedure to set the basic operations:)
.
10.3 Selection
sequence
To do a calibration on a measure or source function, use the
Advanced menu option; see
Section 2.10 (The Advanced
menu options)
:
➤
Configuration
➤
Advanced
➤
Calibration
➤
[Enter PIN:
4321]
➤
Perform
Calibration
➤
Then select a function and start the calibration. This example
shows the Current (measure) function:
Resistance
(CH1)
(measure)
(source)
Standard 0
Ω
resistor
An ohmmeter or an RTD
measurement system with the
specified excitation currents; see
Table 10-11
.
*
Standard resistor (
Ω
): 100, 200, 300,
400, 1k, 2k, 4k
Total uncertainty: 20 ppm
CJ
(CH1)
- Calibrated K type thermocouple
Accuracy: 50 mK for -5 to 28°C (23 to 82.4°F)
- Thermocouple temperature reference unit (0°C)
Accuracy: 30 mK
Pressure
(P1 or P2)
An applicable pressure standard (primary or secondary):
Range 25 mbar/0.36 psi
: total uncertainty of 0.015% reading or better
Ranges > 25 mbar/0.36 psi
: total uncertainty of 0.01% reading or better
* Or an equivalent resistance simulator
Table 10-1:
Calibration equipment (Continued)
Function
Calibration equipment
(ppm = parts per million)
2
The display shows the
applicable instructions to
complete the calibration.
TAP
3
1
Select function
Select range
(if applicable)
Select channel
Summary of Contents for Druck DPI 620-IS
Page 4: ...Page 4 of 98 K0460 Issue 1 ...
Page 5: ...GE Measurement Control Systems Druck DPI 620 IS advanced modular calibrator user manual K0460 ...
Page 12: ...Issue 1 viii About this manual K0460 EN English ...
Page 21: ... EN English K0460 Instrument parts accessories and options 1 5 Issue 1 ...
Page 22: ...Issue 1 1 6 Instrument parts accessories and options K0460 EN English ...
Page 42: ...Issue 1 2 20 Prepare the instrument K0460 EN English ...
Page 60: ...Issue 1 5 2 Instrument communications K0460 EN English ...
Page 64: ...Issue 1 6 4 Datalog operation K0460 EN English ...
Page 78: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Page 94: ...Issue 1 10 16 Calibration procedures K0460 EN English ...
Page 96: ...Issue 1 11 2 General specification K0460 EN English ...
Page 97: ...Customer service Visit our web site www gesensinginspection com ...
Page 98: ......