Issue 1
3-8 Electrical operations
K0460 - [EN] English
3.
If necessary change the
Settings
and continue with the
measure operation.
Values to set
:
TC Type (Set the applicable TC).
CJ compensation (Mode: Automatic/Manual).
Automatic
uses the internal cold junction. Use
Manual
mode for an
external cold junction.
CJ Value. For
Manual
mode, set an applicable value. The
value is not used in
Automatic
mode.
Example B
1.
Set the applicable software options; see
Section 3.2.1
(Procedure overview)
.
2.
Complete the electrical connections.
3.
If necessary, change the
Source Settings
and continue with
source operation.
Values to set
:
TC Type (Set the applicable TC).
CJ compensation (Mode: Automatic/Manual).
Automatic
mode
uses the internal cold junction.
Manual
mode uses an
external cold junction.
CJ Value. For
Manual
mode, set an applicable value. The
value is not used in
Automatic
mode.
4.
Automation
: Set the applicable output value; see
Section 2.9.2
.
3.2.9 Example procedure: Switch test
CH1, P1, P2 functions use the CH2 switch connections; CH2
functions use the CH1 switch connections.
Switch operation
When setting the
Switch Test
utility on one channel, the
software automatically sets-up the other channel for the switch
connections.
Note:
If there is a measure or source function on the switch
connection channel, it is automatically disabled with the screen
message “... Function Disabled”.
An attempt to set a measure or source function on the switch
connection channel, causes a screen message “... Function not
set”.
Example
This example shows a thermocouple switch:
Summary of Contents for Druck DPI 620-IS
Page 4: ...Page 4 of 98 K0460 Issue 1 ...
Page 5: ...GE Measurement Control Systems Druck DPI 620 IS advanced modular calibrator user manual K0460 ...
Page 12: ...Issue 1 viii About this manual K0460 EN English ...
Page 21: ... EN English K0460 Instrument parts accessories and options 1 5 Issue 1 ...
Page 22: ...Issue 1 1 6 Instrument parts accessories and options K0460 EN English ...
Page 42: ...Issue 1 2 20 Prepare the instrument K0460 EN English ...
Page 60: ...Issue 1 5 2 Instrument communications K0460 EN English ...
Page 64: ...Issue 1 6 4 Datalog operation K0460 EN English ...
Page 78: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Page 94: ...Issue 1 10 16 Calibration procedures K0460 EN English ...
Page 96: ...Issue 1 11 2 General specification K0460 EN English ...
Page 97: ...Customer service Visit our web site www gesensinginspection com ...
Page 98: ......