Issue 1
8-8 HART
®
device operations
K0460 - [EN] English
2.
Set-up the
HART
®
device, if loop power required, use CH2 for
current (24V)
.
If the loop is already powered and there is an external HART
®
resistor, turn off CH2. Connect CH2 mA+ and mA- connections
to the HART
®
resistor, see
Section 1.3
.
If required, turn on the internal HART
®
resistor. This can be
selected by pressing
icon.
Press the HART
®
button and the application starts.
3.
Complete the electrical connections and continue with the
HART
®
menu operation; see
Section 8.4
.
4.
Complete the electrical connections and continue with the
HART
®
menu operation; see
Section 8.4
.
8.3.2 External loop power
In this example, there is an external power supply.
Set the functions for channel 2 and HART:
5.
Set the
HART
®
Resistor.
Measure current on
channel 2 (CH2)
Range: ±55 mA
HART Function is set
250
Ω
resistor is set
e
1
a
b
d
c
Set HART
Section 2.8.4
Set Current; Measure
Section 2.8.1
Home display
T
Summary of Contents for Druck DPI 620-IS
Page 4: ...Page 4 of 98 K0460 Issue 1 ...
Page 5: ...GE Measurement Control Systems Druck DPI 620 IS advanced modular calibrator user manual K0460 ...
Page 12: ...Issue 1 viii About this manual K0460 EN English ...
Page 21: ... EN English K0460 Instrument parts accessories and options 1 5 Issue 1 ...
Page 22: ...Issue 1 1 6 Instrument parts accessories and options K0460 EN English ...
Page 42: ...Issue 1 2 20 Prepare the instrument K0460 EN English ...
Page 60: ...Issue 1 5 2 Instrument communications K0460 EN English ...
Page 64: ...Issue 1 6 4 Datalog operation K0460 EN English ...
Page 78: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Page 94: ...Issue 1 10 16 Calibration procedures K0460 EN English ...
Page 96: ...Issue 1 11 2 General specification K0460 EN English ...
Page 97: ...Customer service Visit our web site www gesensinginspection com ...
Page 98: ......