3.5
Set the Pressure Module to Zero
SETTINGS
>> ZERO >> ZERO
Use this option to write a new zero pressure value to the
pressure module in use. The sensor adjustment is
permitted if it obeys the condition that follows:
• Adjustment ≤10% FS positive pressure value (for the
Sensor).
Note
: To make a temporary adjustment for zero, use the
Tare function.