User Manual
Druck DPI620G-IS
29 July 2016
116M5464 Revision -
Page 69 of 248
1.
Set the applicable software options.
2.
Complete the electrical connections.
3.
If necessary change the thermocouple type.
SETTINGS
>> TC TYPE
4.
Set CJ Compensation mode.
SETTINGS
>> MANUAL CJ COMPENSATION
If Manual CJ compensation is not selected, the
internal cold junction is used to calculate the
thermocouple value.
If using an external cold junction, select check box on
MANUAL CJ COMPENSATION
and enter value for cold
junction temperature.
Figure 2-28 K-Type Thermocouple Source CH1 (Range
-270 to 1372 °C)