User Manual
Druck DPI620G-IS
29 July 2016
116M5464 Revision -
Page 52 of 248
2.5
Measure and Source Operations
After setting the required measure and source functions on
the display, additional features can be set for each
displayed function
These features are selected by entering the function’s
expanded view and selecting
Measure Function PROCESSES
Processes available depend on the function and include:
a)
Tare:
Used to set a temporary value for zero. This
makes an adjustment to all subsequent readings on the
display
b)
Alarm:
Used to indicate whena limit has been exceeded
c)
Filter:
Sets the Band and Time Constant for a low pass
filter
d)
Flow:
When selected, the square root of the measured
value is displayed.
e)
Scaling:
Absolute values are scaled
Source Function SETTINGS
Settings available depend on the function and include:
a)
AUTOMATION:
allows values on the display to be set
and changed by the following PROCESSES:
Nudge:
value adjustment by up/down arrows
Span Check:
Alternately apply specified HIGH & LOW
values under time control or up/down arrow control.