Function
Calibration Equipment
(ppm = parts per million)
Cold Junction
(CH1)
Calibrated K type thermocouple
Accuracy: 50 mK for -5 to 28°C (23 to 82.4°F)
Thermocouple temperature reference unit
(0°C).
Accuracy: 30 mK
Pressure
(P1 or P2)
DPI620G-IS Module Carrier MC 620G-IS or
DPI620G-IS Pressure Base PV 62XG-IS:
Range 25 mbar/0.36 psi
: total uncertainty of
0.015% reading or better
Ranges > 25 mbar/0.36 psi
: total uncertainty
of 0.01% reading or better
IDOS
UPM only. Refer to the user manual for the
IDOS UPM.
Before starting the calibration, check that the time and date on