Example Procedure: Measure or Source Current, CH2
2.6.2
Note:
CH2 can be configured to measure or source
current using internal or external loop power. For
examples of internal loop power, refer to
Section 2.7
Figure 2-14 shows CH2 set-up to measure or source a
current with external loop power.
Figure 2-14 Measure Current on CH2, Range ± 55 mA
1.
Set the applicable software option: CH2,
Measure/Source, Current, mA.
2.
Complete the electrical connections and continue
with the measure or source operation.
3.
Source only (Automation): Set the applicable
output value.