User Manual
Druck DPI620G-IS
29 July 2016
116M5464 Revision -
Page 223 of 248
10.6
Calibration Procedures (CH1): DC mV/Volts (source)
Note:
When performing a calibration of both the measure
and
source functions, the measure calibration
should be performed first.
Perform the calibration as follows:
1.
Connect the applicable calibration equipment specified
in Table 10-1; for example see
Section 2.9
2.
Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3.
Use the calibration menu (Ref:
Section 10.2
) to do a
two-point calibration (Zero and +FS) for the applicable
range:
mV (source) ranges
Volts (source) ranges
2000 mV
12 V
4.
Check the calibration is correct:
•
Select the applicable Millivolts or Voltage (source)
function (ref
Section 2.9
•
Apply the input values that are applicable to the
calibration (Ref: Table 10-6
mV: 0, 100, 195, 1000, 1995
Volts (V): 0, 3, 6, 9, 12.
•
Check the error is within the specified limits (Ref:
Table 10-6