User Manual
Druck DPI620G-IS
29 July 2016
116M5464 Revision -
Page 217 of 248
10.2
Selection Sequence
To perform a calibration on a measure or source function,
use the advanced menu option:
1.
Select:
DASHBOARD >>
ADVANCED
2.
Enter the calibration PIN: 4321
3.
Select the button.
4.
Select:
PERFORM CALIBRATION
Note:
At this point, if the instrument has previously been
calibrated, a window will appear allowing the existing
calibration data to be saved (it can later be retrieved via
the ‘ADVANCED’ menu). Choose the preferred option and
continue:
5.
Select channel
6.
Select function
7.
Select range (if applicable)
8.
Follow the on-screen instructions
When the calibration is complete, set the next calibration
date.