User Manual
Druck DPI620G-IS
29 July 2016
116M5464 Revision -
Page 224 of 248
Table 10-6 millivolts (source) error limits
Source
(mV)
Calibrator
uncertainty
(mV)
Permitted
DPI620G-IS
error (mV)
0
0.0001
±0.0060
100
0.00046
±0.0090
195
0.0009
±0.0119
1000
0.003
±0.0750
1995
0.006
±0.0899
Table 10-7 Voltage (source) error limits
Source
(V)
Calibrator
uncertainty
(V)
Permitted
DPI620G-IS
error (V)
0
0.000004
±0.00048
3
0.000019
±0.00060
6
0.000034
±0.00072
9
0.000049
±0.00084
12
0.000064
±0.00096