Revsion A
ProCare
Partient Monitor Service Manual
3-5
2009381-001
Principles of Operation:
3.3 Functional Description
3.2.3 Temperature (Model 200 and 400)
The
ProCare
Monitor uses Alaris Turbo Temp technology to measure patient
temperature. The Turbo Temp probe contains a heating element that pre-
heats the probe to reduce determination time. The heating function is
controlled by the Main Board. The Turbo Temp probe also contains a
thermistor that indicates the temperature. When the probe is attached to the
temperature connector and patient, the signal generated by the thermistor is
routed to the Main Board. The Main Board converts the thermistor signal
along with status information (i.e ORAL or RECTAL probe indicators) to a
DIGITAL signal. The Main Board then processes the DIGITAL signal and
displays the patient temperature on the UI Board and printer in Celsius or
Fahrenheit.
3.2.4 Host Communication Port
The Host Comm Port is used to interface the
ProCare
Monitor with other
electronic devices (a central nurse's station or remote alarm device.) Signals
can be sent to the
ProCare
Monitor to initiate blood pressure determinations
and other functions. Patient data can also be retrieved through this port. For
further information, reference the Dinamap
ProCare
Series Host
Communication manual.
3.3 Functional Description
The following paragraphs provide the functional interface relationship. The
ProCare
Monitor contains a number of electrical & electro-mechanical
assemblies. These assemblies are:
* Main Board PWA
* User Interface (UI) Board PWA
* SPO2 PWA (optional)
* Printer (optional)
* Pneumatic Valve/Manifold (PVM)
* Optical Switch (optional)
Summary of Contents for DINAMAP ProCare 100
Page 1: ......
Page 4: ...DINAMAP ProCare Monitor Service Manual 2009381 001 Revision A ...
Page 9: ...ProCare Patient Monitor Service Manual Revision A 2009381 001 Table of Contents ...
Page 10: ...Revision A ProCare Patient Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Page 11: ...1 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 ...
Page 20: ...Revsion A ProCare Partient Monitor Service Manual 2 1 2009381 001 2 Product Description ...
Page 21: ...2 2 ProCare Partient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 36: ...Revision A ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Principles of Operation ...
Page 37: ...3 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 46: ...Revision A ProCare Patient Monitor Service Manual 4 1 2009381 001 4 Calibration Maintenance ...
Page 47: ...4 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 83: ...PNEUMATIC TUBING ROUTING Exploded Assembly Drawing ProCare Monitor front case page 5 3 4 or ...
Page 96: ...2009381 001 A ...