2-12
ProCare
Partient Monitor Service Manual
Revision A
2009381-001
Product Description:
2.6. Specifications
2.6. Specifications
0 0 8 6
This product conforms with the essential requirements of the Medical Device Directive.
Accessories without the CE Mark are not guaranteed to meet the Essential requirements of
the Medical Device Directive.
IPX1
The ProCare Monitor is protected against vertically falling drops of water and conforms to
the IEC-529 standard at level
IPX1.
Vertically falling drops shall have no harmful effects to
the monitor.
Power Requirements
MAINS
Protection against electrical shock - Class II
AC INPUT VOLTAGE
120 VAC/60 Hz 24W
Alternate Sources
Protection against electrical shock - Class II
DC INPUT VOLTAGE
24 VDC (nominal), at 1A
The AC Mains adapter contains a nonresettable and nonreplaceable fuse.
Battery
6 volt, 3.3 amp-hours. Protected by internal auto-resetting fuse and thermal protection.
Minimum Operation Time
2 hrs (5 min cycle with adult cuff at 25° C, SpO2 active at 60 bpm, temp in monitor mode,
printout of current values every 5 minutes) from full charge.
Fuses
The Monitor contains four fuses. The fuses are auto-resettable and mounted within the
Monitor. The fuses protect the low voltage DC input, the battery, the remote alarm output
and the +5V output on the host port connector.
Environmental
Operating Temperature
+5° C to + 40° C (-41 F to +104 F)
Storage Temperature
-20° C to +50° C (-4 F to +122 F)
Operating Atmospheric
Pressure Range
700 to 1060 hecto Pascal
Humidity Range
0% to 95% non-condensing
Radio Frequency
Complies with IEC Publication 601-1-2 (April 1993) Medical Electrical Equipment,
Electromagnetic Compatibility Requirements and Tests, and CISPR 11 (Group1, Class A)
for radiated and conducted emissions.
Mechanical
Dimensions
Height:
9.8 in (25.0 cm)
Width:
9.8 in (24.8 cm)
Depth:
6.9 in (17.5 cm)
Weight including battery
7.8 lb (3.5 Kg)
Mountings
Self Supporting on rubber feet.
Portability
Carried by built-in handle
Summary of Contents for DINAMAP ProCare 100
Page 1: ......
Page 4: ...DINAMAP ProCare Monitor Service Manual 2009381 001 Revision A ...
Page 9: ...ProCare Patient Monitor Service Manual Revision A 2009381 001 Table of Contents ...
Page 10: ...Revision A ProCare Patient Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Page 11: ...1 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 ...
Page 20: ...Revsion A ProCare Partient Monitor Service Manual 2 1 2009381 001 2 Product Description ...
Page 21: ...2 2 ProCare Partient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 36: ...Revision A ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Principles of Operation ...
Page 37: ...3 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 46: ...Revision A ProCare Patient Monitor Service Manual 4 1 2009381 001 4 Calibration Maintenance ...
Page 47: ...4 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 83: ...PNEUMATIC TUBING ROUTING Exploded Assembly Drawing ProCare Monitor front case page 5 3 4 or ...
Page 96: ...2009381 001 A ...