Revision A
ProCare
Partient Monitor Service Manual
2-15
2009381-001
Product Description:
2.6. Specifications
* The Masimo SET
®
SpO
2
parameter with LNOP-Adt sensors has been
validated for no motion accuracy in human blood studies on healthy adult
volunteers in induced hypoxia studies in the range of 70-100% SpO
2
against
a laboratory co-oximeter and ECG monitor. This variation equals plus or
minus one standard deviation. Plus or minus one standard deviation
encompasses 68% of the population.
**The Masimo SET
®
SpO
2
parameter with LNOP-Adt sensors has been
validated for motion accuracy in human blood studies on healthy adult
volunteers in induced hypoxia studies while performing rubbing and tapping
motions at 2 to 4 Hz at an amplitude of 1 to 2 cm and a nonrepetitive motion
between 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in
the range of 70-100% SpO2 against a laboratory co-oximeter and ECG
monitor. This variation equals plus or minus one standard deviation. Plus or
minus one standard deviation encompasses 68% of the population.
†The Masimo SET
®
SpO
2
parameter with LNOP-Neo Pt sensors has been
validated for neonatal motion accuracy in human blood studies on neonates
while moving the neonate’s foot at 2 to 4 cm against a laboratory co-
oximeter and ECG monitor. This variation equals plus or minus, one
standard deviation.Plus or minus one standard deviation encompasses 68%
of the population.
‡The Masimo SET
®
SpO
2
parameter with has been validated for low
perfusion accuracy in bench top testing against a Biotek Index 2 stimulator
and Masimo’s simulator with signal strengths of greater than 0.02% and a %
transmission of greater than 5% for saturations ranging from 70 to 100%.
This variation equals plus or minus, one standard deviation.Plus or minus
one standard deviation encompasses 68% of the population.
Accuracy and Motion Tolerance
Without Motion - Adult/Ped*
70 to 100% ±2 digits
Without Motion - Neonate*
70 to 100% ±3 digits
With Motion - Adult/Ped/Neo**†
70 to 100% ±3 digits
Low Perfusion‡
70 to 100% ±2 digits
0 to 69% unspecified
Pulse Rate
Without Motion
25 to 240 beats/min ±3 digits
With Motion
normal physiologic range
25 to 240 beats/min ±5 digits
SpO
2
- Masimo Oximetry
Summary of Contents for DINAMAP ProCare 100
Page 1: ......
Page 4: ...DINAMAP ProCare Monitor Service Manual 2009381 001 Revision A ...
Page 9: ...ProCare Patient Monitor Service Manual Revision A 2009381 001 Table of Contents ...
Page 10: ...Revision A ProCare Patient Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Page 11: ...1 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 ...
Page 20: ...Revsion A ProCare Partient Monitor Service Manual 2 1 2009381 001 2 Product Description ...
Page 21: ...2 2 ProCare Partient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 36: ...Revision A ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Principles of Operation ...
Page 37: ...3 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 46: ...Revision A ProCare Patient Monitor Service Manual 4 1 2009381 001 4 Calibration Maintenance ...
Page 47: ...4 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 83: ...PNEUMATIC TUBING ROUTING Exploded Assembly Drawing ProCare Monitor front case page 5 3 4 or ...
Page 96: ...2009381 001 A ...