![GE DINAMAP ProCare 100 Service Manual Download Page 65](http://html.mh-extra.com/html/ge/dinamap-procare-100/dinamap-procare-100_service-manual_128373065.webp)
4-20
ProCare
Patient Monitor Service Manual
Revision A
2009381-001
Calibration & Maintenance:
4.7 Alarm Code Interpretation
Temperature / Printer / Miscellaneous
E23
SpO
2
Sensor off finger
No SpO
2
signal, check or reposition the sensor
E25
SpO
2
No Signal
No or very low SpO
2
signal, reposition sensor
Alarm
Definition
Possible Cause
E63
Temp Disconnected or
Wrong Probe Type
Incorrect type of temperature probe: use
TurboTemp-type temperature probe
E61
Temp Probe Broken
Bad temperature probe
Temp probe not properly connected
E66
Temp probe Too Hot
Bad temperature probe
Verify monitor with known good accessories
E10
Printer No Paper
Printer is out of paper
Printer door open
Printer may be defective
E11
Printer Too Hot
E12
Recorder cannot print
Main battery voltage is too low to operate
E00
Memory Lost
Refer Monitor to Customer Service
900-
999
Internal Memory Errors
Refer Monitor to Customer Service
Summary of Contents for DINAMAP ProCare 100
Page 1: ......
Page 4: ...DINAMAP ProCare Monitor Service Manual 2009381 001 Revision A ...
Page 9: ...ProCare Patient Monitor Service Manual Revision A 2009381 001 Table of Contents ...
Page 10: ...Revision A ProCare Patient Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Page 11: ...1 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 ...
Page 20: ...Revsion A ProCare Partient Monitor Service Manual 2 1 2009381 001 2 Product Description ...
Page 21: ...2 2 ProCare Partient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 36: ...Revision A ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Principles of Operation ...
Page 37: ...3 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 46: ...Revision A ProCare Patient Monitor Service Manual 4 1 2009381 001 4 Calibration Maintenance ...
Page 47: ...4 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 83: ...PNEUMATIC TUBING ROUTING Exploded Assembly Drawing ProCare Monitor front case page 5 3 4 or ...
Page 96: ...2009381 001 A ...