Revision A
ProCare
Patient Monitor Service Manual
4-9
2009381-001
Calibration & Maintenance:
4.5 Safety Testing
4.5 Safety Testing
To adequately test the safety and integrity of the
ProCare
Monitor, the
following test equipment is recommended:
•
12VDC Power Supply
•
DMM (Fluke 8842 or equivalent)
•
NIBP Analyzer (DNI Nevada "Cufflink" or Equivalent)
•
Adult BP Cuff, Neonate BP Cuff, hose, inflation bulb, and mandrel
•
Inflation bulb and associated tubing
•
Calibration Kit, p/n 320246 available through GE Medical Systems
•
SPO
2
Simulator (for appropriate SPO
2
type if SPO
2
is installed)
•
SPO
2
Cable (for appropriate SPO
2
type, if SPO
2
is installed)
•
TE 1811 Temperature Probe Simulator (if TEMP is installed.) The
Temperature Simulator for the Alaris System is available from Alaris Medical
Systems, Inc. (619) 458-7000
•
Printer Paper (if PRINTER is installed)
CAUTION! Calibration equipment should always be kept dry and free of
particulate matter. Moisture or foreign substances introduced to the
pneumatic system will likely cause damage to the monitor and/or the
accessories.
Complete the Test Record (Appendix A) as tests are performed.
Note:
This test is written so that a knowledgeable technician who is familiar
with the
ProCare
monitor and the test equipment and will be able to follow
the test procedure.
Note:
To enter service mode press and hold the CYCLE buttons while
pressing the ON/OFF button for 3 seconds.
Notes on Electrical Safety Testing of the
ProCare
Monitor:
The DINAMAP
®
ProCare
Monitor is designed and tested to meet electrical
safety standard IEC 601-1. Requirements in this standard parallel
requirements in NFPA99 relating to electrical safety. This product meets the
requirements of NFPA99 section 7-5.1.2.2. (grounding of appliances)
through the use of double insulation in the external power supply module
(“Power Brick”.)
The absence of a ground connection on the power brick obviates the need
for any ground resistance test, such as that described in NFPA99 section 7-
5.1.3.2. (Some non-USA Power bricks are equipped with a non-conductive
Summary of Contents for DINAMAP ProCare 100
Page 1: ......
Page 4: ...DINAMAP ProCare Monitor Service Manual 2009381 001 Revision A ...
Page 9: ...ProCare Patient Monitor Service Manual Revision A 2009381 001 Table of Contents ...
Page 10: ...Revision A ProCare Patient Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Page 11: ...1 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 ...
Page 20: ...Revsion A ProCare Partient Monitor Service Manual 2 1 2009381 001 2 Product Description ...
Page 21: ...2 2 ProCare Partient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 36: ...Revision A ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Principles of Operation ...
Page 37: ...3 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 46: ...Revision A ProCare Patient Monitor Service Manual 4 1 2009381 001 4 Calibration Maintenance ...
Page 47: ...4 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 83: ...PNEUMATIC TUBING ROUTING Exploded Assembly Drawing ProCare Monitor front case page 5 3 4 or ...
Page 96: ...2009381 001 A ...