Revision A
ProCare
Patient Monitor Service Manual
4-11
2009381-001
Calibration & Maintenance:
4.6 ProCare Patient Monitor Parameter Tests
4.6
ProCare
Patient Monitor Parameter Tests
Complete the Test Record (Appendix A) as tests are performed.
Note: This test is written so that a technician who is familiar with the
ProCare
monitor and the test equipment will be able to successfully
complete the test procedures. To enter service mode press and hold
the CYCLE button while pressing the ON/OFF button for 3 seconds.
Saving Changes into Flash Memory:
Modifications made to the Monitor’s
settings/configuration must be saved prior to placing the Monitor into
service. The Monitor must be in service mode for any changes to be saved.
Events that require you to manually save to the Monitor’s memory include
calibration attempts and enabling/disabling the Monitor’s parameters (refer
to section 4.6.4.) It is not necessary to manually save following a pressure
transducer calibration verification or simple calibration check.
The number of times the flash memory can be saved is limited. The fatal
alarm error
975
may be issued after this process is performed more than 90
times. Following issuance of this fatal alarm, the monitor must be returned to
the GE Medical Systems
Information Technologies
Service Center to have
the flash memory reset so that future saves can be performed.
4.6.1 SETUP
1. Connect your NIBP Analyzer to the
ProCare
Monitor.
2. ‘T’ an inflation bulb into the pneumatic setup
3. Consult the following diagram for pneumatic setup guidelines.
Summary of Contents for DINAMAP ProCare 100
Page 1: ......
Page 4: ...DINAMAP ProCare Monitor Service Manual 2009381 001 Revision A ...
Page 9: ...ProCare Patient Monitor Service Manual Revision A 2009381 001 Table of Contents ...
Page 10: ...Revision A ProCare Patient Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Page 11: ...1 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 ...
Page 20: ...Revsion A ProCare Partient Monitor Service Manual 2 1 2009381 001 2 Product Description ...
Page 21: ...2 2 ProCare Partient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 36: ...Revision A ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Principles of Operation ...
Page 37: ...3 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 46: ...Revision A ProCare Patient Monitor Service Manual 4 1 2009381 001 4 Calibration Maintenance ...
Page 47: ...4 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 83: ...PNEUMATIC TUBING ROUTING Exploded Assembly Drawing ProCare Monitor front case page 5 3 4 or ...
Page 96: ...2009381 001 A ...