Revision A
ProCare
Patient Monitor Service Manual
4-13
2009381-001
Calibration & Maintenance:
4.6 ProCare Patient Monitor Parameter Tests
12. Record and verify the pressure reading on the top LED display
(SYSTOLIC)
13. Record and verify the pressure reading on the bottom LED display
(DIASTOLIC)
14. Use the valve on the bulb to reduce pressure to 50mmHg
15. Record and verify the pressure reading on the top LED display
(SYSTOLIC)
16. Record and verify the pressure reading on the bottom LED display
(DIASTOLIC)
If any of the tests fail, continue to section 4.6.4. Otherwise, continue to
section 4.6.5
4.6.4 Pressure Transducer Calibration
Perform only if Pressure Transducer Verification is out of tolerance as
specified in Appendix A.
Note:
To enter service mode press and hold the
CYCLE
button
while
pressing the
ON/OFF
button.
1. Turn the
ProCare
Monitor
ON
and enter Service Mode.
2. The
min
window should display
0
.
3. Open valve on bulb to open pressure system to atmosphere
4. Set NIBP Analyzer to "Manometer" function and press "Zero Pressure"
5. Press
CYCLE
button until the
min
window shows 1.
6. Close valve on bulb and manually inflate pressure to 200 mmHg (using
the CuffLink manometer as reference).
7. Press
MENU
button when pressure reads 200 mmHg to save calibration
setting.
8. Two steps are necessary for burning changes into flash:
a. Press the
CYCLE
button repeatedly until the number
6
is displayed in
the
min
window.
Note: the number of remaining changes to the flash memory available are
displayed in the
MAP/Cuff
window. Following each true calibration, the
Monitor decrements by 1 the number of remaining calibrations available. If
the MAP/Cuff window is displaying a figure of less than 10 and the accuracy
of the Monitor is in doubt, contact General Electric Medical Systems
Information Technologies
Technical Support at
1-877-274-8456.
b. Press and hold the
MENU
button until two beeps are heard (one when
Summary of Contents for DINAMAP ProCare 100
Page 1: ......
Page 4: ...DINAMAP ProCare Monitor Service Manual 2009381 001 Revision A ...
Page 9: ...ProCare Patient Monitor Service Manual Revision A 2009381 001 Table of Contents ...
Page 10: ...Revision A ProCare Patient Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Page 11: ...1 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 ...
Page 20: ...Revsion A ProCare Partient Monitor Service Manual 2 1 2009381 001 2 Product Description ...
Page 21: ...2 2 ProCare Partient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 36: ...Revision A ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Principles of Operation ...
Page 37: ...3 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 46: ...Revision A ProCare Patient Monitor Service Manual 4 1 2009381 001 4 Calibration Maintenance ...
Page 47: ...4 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 83: ...PNEUMATIC TUBING ROUTING Exploded Assembly Drawing ProCare Monitor front case page 5 3 4 or ...
Page 96: ...2009381 001 A ...