Action
Step
Disconnect the tubing from the top of the UV monitor and replace it with a
piece of tubing leading to the waste container.
2
W
Flush the flow cell with 25 to 30 mL distilled water, using a syringe attached
to the Luer-to-SNAP connector.
3
Flush the flow cell about five times with 10 mL detergent solution.
Tip:
Warm the detergent solution to 40°C for enhanced cleaning effect.
4
Allow the detergent solution to remain in the flow cell for about 15 minutes.
5
Flush the flow cell twice with 25 to 30 mL distilled water.
6
Disconnect the Luer-to-SNAP connector and waste tubing from the UV
monitor, and reconnect the system tubing.
7
ÄKTA pilot 600 User Manual 29274559 AA
99
5 Maintenance
5.3 Cleaning procedures
5.3.2 Clean the UV flow cell