The reports are saved in the folder
C:\Program Files (x86)\GE
Healthcare\UNICORN\UNICORN <version number>\Temp
. Each test generates
a text file with the name of the test as filename. An incremented serial number is added
if a file with the same name already exists.
The dialog for selecting and running a performance test includes a tab with
heading System Performance Report (see
). This tab
does not contain the reports for performance tests.
Note:
Test results
The performance test data are displayed in the System Control module as the test pro-
gresses, and are saved in the same way as results from chromatography runs. Examina-
tion of the results can help in assessing the outcome of the tests, particularly for tests
that fail. The illustration below shows the results of a System test, displayed in the System
Control module as the test was run.
You can choose not to save the test results when a test is run (see
This does not affect generation of the text report.
ÄKTA pilot 600 User Manual 29274559 AA
127
6 Performance tests
6.3 How to evaluate performance tests