Action
Step
Set the following parameters:
5
•
Column hardware
•
Bed support
•
Resin
•
Target bed height
Result: Default values are automatically set to provide a complete method.
Select GE verified packing settings to get a packing method for GE resins
verified by GE. If you use a custom resin, the only option is Custom packing
settings.
Tip:
When GE verified packing settings is selected, you can enter the slurry
concentration and click Slurry Recipe to get a slurry recipe calculated to give
your target bed height.
6
Verify the inlets for hydraulic chamber liquid and mobile phase.
Tip:
Click Show Column Connection to find an illustration of the connections to
the column, and read the time estimation.
7
Verify the phase properties for the remaining phases.
8
Save the method.
9
ÄKTA pilot 600 User Manual 29274559 AA