Pressure protection
The UV monitor is sensitive to high pressure, but normally the pressure is low after the
column. To protect sensitive monitors after the column, the post-column pressure has
an alarm level of 0.8 MPa. This alarm is independent of other settings.
UV cell size
The UV flow cell is available in two sizes:
•
2 mm UV flow cell, can be used in all applications
•
5 mm UV flow cell, is recommended for peak volumes larger than 1000 mL
Turning off the lamp
The UV lamp can be turned off manually if not needed during a run, for example during
column or system conditioning or equilibration over long time, cleaning column or system,
or sanitization. It will be turned on automatically for the next run.
Follow the steps below to turn off the lamp:
Action
Step
Click Manual:Execute manual instructions in the System Control module
to open the Manual instructions dialog box.
1
Select Monitors:UV lamp, select Off and click Execute.
2
Cleaning
See
Section 5.3.2 Clean the UV flow cell, on page 98
.