E
NGINE
P
RESSURE
R
ATIO
(EPR)
K0553 Revision D
88
[EN] English
Figure 3- 32: EPR Control Screen (Method2)
3)
Touch the “Aim” value (2). A numbered keypad opens.
4)
Enter the desired “Aim” value and touch the “tick” icon. The numbered keypad closes and
the previously entered value appears in the “IN” display.
5)
Touch the “Aim” value (3). A numbered keypad opens.
6)
Enter the desired “Aim” value and touch the “tick” icon. The numbered keypad closes and
the previously entered value appears in the “OUT” display.
Wait for the controller to achieve the IN and OUT aim values. The EPR has now been set based
on the inlet pressure and outlet pressure.
On completion of the EPR checks, select “Go to ground” and wait for the AT GROUND message
to be displayed, see also section Go to ground. Refit the pitot and static blanking caps.
- EPR Limits
Min Inlet
60 mbar
Max Inlet
1355 mbar
Min Outlet
60 mbar
Max Outlet
2000 mbar
Min (Ratio)
0.1
Summary of Contents for ADTS552F
Page 19: ...Chapter 1 Introduction...
Page 30: ...DUAL HAND TERMINAL EN English 19 K0553 Revision D Figure 1 8 ADTS Touch Slave...
Page 31: ...Chapter 2 Installation...
Page 34: ...PACKING FOR STORAGE OR TRANSPORTATION EN English 24 K0553 Revision D...
Page 40: ...Chapter 3 Operation...
Page 114: ...Chapter 4 Calibration...
Page 123: ...Chapter 5 Maintenance...
Page 132: ...Chapter 6 Testing and Fault Finding...
Page 152: ...Chapter 7 Specification...