ADTS leak check
Before carrying out a leak check, allow at least 15 minutes for the ADTS to warm-up.
Set-up
1)
From
Dashboard
>>
Settings
.
2)
In the
Settings
>>
ADTS Settings
.
3)
Select
ADTS LIMITS
>> then
SELECT LIMITS"
>> select
MAX AERO
.
4)
Return to
ADTS Settings
.
5)
Select
Pressure units
.
6)
Select the “mbar” radio button.
7)
Exit the
Settings
menu and return to the
Dashboard.
8)
Select
PITOT STATIC
.
Pressure leak check
This procedure verifies that the unit is leak tight under positive pressure conditions.
1)
Select
CONTROL
.
2)
Toggle Pressure Aeronautical Units to mbar.
3)
Swipe screen across right to left to select the Rate Timer and set the “WAIT” time to 5
minutes and 0 seconds and touch the “tick” icon.
4)
Set the “TEST” time to 1 minute and 0 seconds and touch the “tick” icon.
5)
Swipe back to the control mode.
6)
Touch the Ps “AIM” value to highlight it.
7)
Use the numbered keypad to input the new value: 1050.00mbar, set rate aim to
500mbar/min and touch the “tick” icon.
8)
Touch the Qc “AIM” value to highlight it.
9)
Use the numbered keypad to input the new value: 860.00mbar, set rate aim to
500mbar/min and touch the “tick” icon.
10) Wait until the Ps and Qc have reached the “Aim” values and stabilized (indications will be
green).
11) Select
MEASURE
.
12) Swipe across to select the Rate Timer.
Summary of Contents for ADTS552F
Page 19: ...Chapter 1 Introduction...
Page 30: ...DUAL HAND TERMINAL EN English 19 K0553 Revision D Figure 1 8 ADTS Touch Slave...
Page 31: ...Chapter 2 Installation...
Page 34: ...PACKING FOR STORAGE OR TRANSPORTATION EN English 24 K0553 Revision D...
Page 40: ...Chapter 3 Operation...
Page 114: ...Chapter 4 Calibration...
Page 123: ...Chapter 5 Maintenance...
Page 132: ...Chapter 6 Testing and Fault Finding...
Page 152: ...Chapter 7 Specification...