C
ALIBRATION
PROCESS
[EN] English
112
K0553 Revision D
Table 4-2:
Calibration requirements ADTS552/553/554F
Table 4-3:
Calibration equipment uncertainty
Suggested two-point calibration adjustment points
Calibration points are nominal pressures to allow for slight differences due to primary standard
weights and actual corrected pressure calculations.
- Ps: two point cal points = 1st point is 1128mbar (FS) followed by 92mbar (suggest to be done in
this order).
- Pt: two point cal points = 1st point 92mbar followed by 1997mbar (FS) (suggest to be done in
this order).
ADTS552/553/554F
Accuracy specifications(k = 2,
95% uncertainty)Includes: NL, H
and R over the ADTS operating
temperature range and
calibration stability over 12
months and calibration
equipment uncertainty stated in
Table 4-3
Test method
Pneumatic pressure
Ps Range:
72 mbar to 1130 mbar absolute
(EALT option 57 mbar)
Accuracy:
Refer to current product data sheet
Compared against calibration
standard
Pt Range:
72 mbar to 1997 mbar absolute
(EALT option 57 mbar)
Accuracy:
Refer to current product data sheet
Equipment Type
Minimum use specifications
Purpose
Calibration standard (absolute)
Range:
35 mbar to 2000 mbar absolute.
Expanded uncertainty (k = 2):
32ppm of r0.007
mbar(0.0032% of r0.70 Pa)
Calibration
All traceable to national standards.
If a calibration standard with higher uncertainty is used, the ADTS accuracy will be degraded and may exceed
the ADTS sales data sheet specification.
Summary of Contents for ADTS552F
Page 19: ...Chapter 1 Introduction...
Page 30: ...DUAL HAND TERMINAL EN English 19 K0553 Revision D Figure 1 8 ADTS Touch Slave...
Page 31: ...Chapter 2 Installation...
Page 34: ...PACKING FOR STORAGE OR TRANSPORTATION EN English 24 K0553 Revision D...
Page 40: ...Chapter 3 Operation...
Page 114: ...Chapter 4 Calibration...
Page 123: ...Chapter 5 Maintenance...
Page 132: ...Chapter 6 Testing and Fault Finding...
Page 152: ...Chapter 7 Specification...