ADTS
LEAK
CHECK
[EN] English
138
K0553 Revision D
13) To start the timer, touch the “play” icon. The timer starts counting down, the elapsed time
(percentage) indicator turns blue and the word “WAIT” is displayed.
14) When the timer reaches 100%, it starts to count down again and the word “TEST” is
displayed. When the timer reaches 100%, it stops, the elapsed time indicator remains blue
and the word “END” is displayed.
15) The display shows the timed rates with a “T” after each value. Check that the Ps, and Qc
rates are less than or equal to ±0.6 mbar/min. If the rates are greater than this value, allow
further thermal stabilization time and re-test.
Vacuum leak check
This procedure verifies that the unit is leak tight under negative pressure conditions.
1)
Select
CONTROL
.
2)
Swipe across to select the Rate Timer and set the “WAIT” time to 5 minutes and 0 seconds
and touch the “tick” icon.
3)
Set the “TEST” time to 1 minute and 0 seconds and touch the “tick” icon.
4)
Swipe back to the control mode.
5)
Touch the Ps “AIM” value to highlight it.
6)
Use the numbered keypad to input the new value: 100.00mbar, set rate aim to
500mbar/min and touch the “tick” icon.
7)
Touch the Qc “AIM” value to highlight it.
8)
Use the numbered keypad to input the new value: 0.00mbar, set rate aim to 500mbar/min
and touch the “tick” icon.
9)
Wait until the Ps and Qc have reached the “Aim” values and stabilized (indications will be
green).
10) Select
MEASURE
.
11) Swipe across to select the Rate Timer.
12) To start the timer, touch the “play” icon. The timer starts counting down, the elapsed time
(percentage) indicator turns blue and the word “WAIT” is displayed.
13) When the timer reaches 100%, it starts to count down again and the word “TEST” is
displayed. When the timer reaches 100%, it stops, the elapsed time indicator remains blue
and the word “END” is displayed.
14) The display shows the timed rates with a “T” after each value. Check that the Ps, and Qc
rates are less than or equal to ±0.6 mbar/min. If the rates are greater than this value, allow
further thermal stabilization time and re-test.
If repeated failures occur, return the unit to GE or a GE authorised repair depot.
Summary of Contents for ADTS552F
Page 19: ...Chapter 1 Introduction...
Page 30: ...DUAL HAND TERMINAL EN English 19 K0553 Revision D Figure 1 8 ADTS Touch Slave...
Page 31: ...Chapter 2 Installation...
Page 34: ...PACKING FOR STORAGE OR TRANSPORTATION EN English 24 K0553 Revision D...
Page 40: ...Chapter 3 Operation...
Page 114: ...Chapter 4 Calibration...
Page 123: ...Chapter 5 Maintenance...
Page 132: ...Chapter 6 Testing and Fault Finding...
Page 152: ...Chapter 7 Specification...